Delta 3300
Energy Output Measurement:
Maximum Energy: | 800 J |
Display Resolution: |
Based on Auto Range |
Auto Range: |
2 - 9.99 J, 10-99.9 J, 100-800 J |
Resolution: | 0.01 J ( 2 -9.99), 0.1 J (10-99.9), 1 J (100-500) |
Accuracy: |
±1 % of reading ± 0.1 J (2-9.9), ± 1J (10-99), ± 2 % (100-500) ± 1J |
Load Resistance: | 50 ohms ±1 %; non-inductive (< 1 μH) |
Sampling Time: |
20 uS |
Measurement Time Window: |
60 ms |
Absolute Max Peak Voltage: |
5000 V |
Absolute Max Peak Current: |
100 A |
Pulse Width: |
60 ms |
High-Level Output:
High Range: |
100:1 amplitude attenuation |
Low Range: |
200:1 amplitude attenuation |
Waveform Storage / Playback: |
Discharge can be viewed via ECG outputs and paddles |
Sync Time Measurements:
Timing Window: | Starts: 40 ms at each R wave peak |
Test Waveforms: | All Waveform simulations available |
Delay Time Accuracy: |
±1 ms |
Charge Time Measurement: | 30 to 60 in steps of 15 seconds |
TRANSCUTANEOUS PACEMAKER ANALYZER
Test Load Range: | 50 -1100 Ohms in steps of 100 Ohms. |
Accuracy: | 50 - 500 ohm ±1 %, 500 - 1100 ohm ±1.5 % |
Pulse Measurements:
Amplitude: | 4 to 400 mA (50 ohm load) |
Accuracy: |
±1 % or ± 1 mA |
Max. amplitude: |
400 mA (50 Ohm Load) |
Rate: |
30 to 800 ppm |
Accuracy: |
±1 % or 2 ppm |
Pulse Width: |
0.5 to 50 ms |
Accuracy: | ±1 % or ±0.3 ms |
Demand Sensitivity Test:
Waveforms: | QRS, Square(SQR), Triangle(TRI), and Haversine (SSQ) |
Waveform width: |
40, 50, 60, 70, 80, 90, 100, 200, 300, 400 ms |
Pacer Rate: |
30 to 120 ppm |
ECG Output:
Amplitude: |
User Selectable 0.1 - 5 mV |
Pacer Input (Load dependent):
Amplitude (50 ohm): |
0 - 10 mV |
Resolution (50 ohm): |
25 μV |
Amplitude: (500 ohm): |
0 - 100 mV |
Resolution: (500 ohm): |
1 mV |
Accuracy: |
±2 % |
Immunity Test:
Interference Signal: | 50/60 Hz |
ECG signal Amplitude: |
0.5 to 5.0 mV |
Refractory Period Measurement: |
20 to 500 ms (both Pacing and Sensing) |
Accuracy: |
±5 ms |
PATIENT SIMULATOR
ECG General:
Lead Configuration: |
12-lead simulation RL, RA, LA, LL, V1-6 |
Output Impedance: |
RL to All leads 1000 Ohms, V Leads: 1000 ohms to RL |
High Level Output: |
0.5 V/mV of low level (lead II) |
Rates: |
30, 40, 50, 60, 70, 80, 90, 100, 110, 120,130, 140, 150, 160, 170, 180, 190, 200, 220, 240, 260, 280, 300 & 310 BPM |
Accuracy: | ±1 % |
Amplitude: | User Selectable 0.1 - 5 mV (Lead II) |
Accuracy: | ±2 % @ Lead II (SQR) |
Arrhythmia Selections: |
Ventricular Fibrillation Coarse (VFBC) |
Performance Waveforms: | Sine, Square, Triangle |
Rate: |
0.2, 0.3, 0.4, 0.5, 1, 2, 3, 4, 5, 6, 7, 8, 9, 10, 15, 20, 25, 30, 35, 40, 45, 50, 55, 60 Hz |
Amplitude: |
0.1, 0.25, 0.50, 1, 1.5, 2.0, 2.5, 3.0, 3.5, 4.0, 4.5, 5.0 mV |
Amplitude Accuracy: |
±2 % (Square Wave -Lead II 1.0 mV) |
Pacer Pulse:
Pulse Width: |
10mS |
Amplitude: |
1.0 mV (Lead II) |
Accuracy: | ± 2 % |
GENERAL INFORMATION
Temperature Range:
Operating: |
+15°C to +35°C |
Storage: |
0°C to +50°C |
Display:
Type: |
LCD with LED backlight |
Data Input/Outputs: | RS-232C (1) for Computer control |
Power: | 7.4 Volt Li-Ion Battery 100-230 VAC input, 850 mA DC Charger |
Housing: |
High impact ABS Plastic case |
Dimensions (D x W x H): | 10.25 x 6.25 x 2.5 Inch, 26 X 16 x 6.4 Cm |
Weight: | 1.5 kg / 3.3 lbs |
DELTA 3300
Defibrillator / Transcutaneous Pacemaker Analyzer
Keep your vital life-support equipment in prime condition with the DELTA 3300 Defibrillator / Transcutaneous Pacemaker Analyzer. This purposefully engineered analyzer safeguards the operation of your defibrillators and external (transcutaneous) pacemakers.
The high adaptability of the device makes it capable of defibrillator testing, transcutaneous pacemaker analysis, and 12-Lead ECG/arrhythmia simulations. You can also routinely track your devices’ performance overtime. Test results can be conveniently saved in the internal memory, uploaded to your PC, and printed using ProComPrint software. (CD included with the unit.)
All that capability is built into one durable, yet lightweight, easy-to-use device!
FEATURES
- Compact, lightweight and easy-to-use
- Tests all defibrillators including AEDS
- Measures monophasic and Bi-Phasic waveforms
- Energy and cardioversion measurement
- Pacemaker loads from 50 to 1100 Ohm
- Large LCD Display with Backlight
- Pacer amplitude, rate, pulse width
- Waveform storage and playback
- Li-Ion rechargeable battery
- Save, upload, and print test results using Netech ProComPrint software
- Hard carrying case included
- 2 Year Warranty
- Product Code: 660
- Availability: In Stock